BrachyView: Proof-of-principle of a novel in-body gamma camera for low dose-rate prostate brachytherapy Journal Article


Authors: Petasecca, M.; Loo, K. J.; Safavi-Naeini, M.; Han, Z.; Metcalfe, P. E.; Meikle, S.; Pospisil, S.; Jakubek, J.; Bucci, J. A.; Zaider, M.; Lerch, M. L. F.; Qi, Y.; Rosenfeld, A. B.
Article Title: BrachyView: Proof-of-principle of a novel in-body gamma camera for low dose-rate prostate brachytherapy
Abstract: Purpose: The conformity of the achieved dose distribution to the treatment plan strongly correlates with the accuracy of seed implantation in a prostate brachytherapy treatment procedure. Incorrect seed placement leads to both short and long term complications, including urethral and rectal toxicity. The authors present BrachyView, a novel concept of a fast intraoperative treatment planning system, to provide real-time seed placement information based on in-body gamma camera data. BrachyView combines the high spatial resolution of a pixellated silicon detector (Medipix2) with the volumetric information acquired by a transrectal ultrasound (TRUS). The two systems will be embedded in the same probe so as to provide anatomically correct seed positions for intraoperative planning and postimplant dosimetry. Dosimetric calculations are based on the TG-43 method using the real position of the seeds. The purpose of this paper is to demonstrate the feasibility of BrachyView using the Medipix2 pixel detector and a pinhole collimator to reconstruct the real-time 3D position of low dose-rate brachytherapy seeds in a phantom. Methods: BrachyView incorporates three Medipix2 detectors coupled to a multipinhole collimator. Three-dimensionally triangulated seed positions from multiple planar images are used to determine the seed placement in a PMMA prostate phantom in real time. MATLAB codes were used to test the reconstruction method and to optimize the device geometry. Results: The results presented in this paper show a 3D position reconstruction accuracy of the seed in the range of 0.5-3 mm for a 10-60 mm seed-to-detector distance interval (Z direction), respectively. The BrachyView system also demonstrates a spatial resolution of 0.25 mm in the XY plane for sources at 10 mm distance from Medipix2 detector plane, comparable to the theoretical value calculated for an equivalent gamma camera arrangement. The authors successfully demonstrated the capability of BrachyView for real-time imaging (using a 3 s data acquisition time) of different brachytherapy seed configurations (with an activity of 0.05 U) throughout a 60 × 60 × 60 mm3 Perspex prostate phantom. Conclusions: The newly developed miniature gamma camera component of BrachyView, with its high spatial resolution and real time capability, allows accurate 3D localization of seeds in a prostate phantom. Combination of the gamma camera with TRUS in a single probe will complete the BrachyView system. © 2013 American Association of Physicists in Medicine.
Keywords: treatment outcome; treatment planning; cancer radiotherapy; radiation dose; methodology; sensitivity and specificity; reproducibility; reproducibility of results; radiotherapy dosage; prostate cancer; prostatic neoplasms; feasibility study; iodine 125; pilot study; pilot projects; dosimetry; image quality; three dimensional imaging; prostate tumor; scintiscanning; gamma cameras; brachytherapy; collimator; equipment design; phantom; phantoms, imaging; radiation dose distribution; transrectal ultrasonography; equipment failure; radionuclide imaging; image reconstruction; equipment; equipment failure analysis; mechanical probe; image guided radiotherapy; qa; fourier transformation; gamma camera; radiotherapy, image-guided; 3d seed position reconstruction; brachyview; intraoperative planning system; radioactive seed; measurement accuracy; radiation dose reduction; triangulation
Journal Title: Medical Physics
Volume: 40
Issue: 4
ISSN: 0094-2405
Publisher: American Association of Physicists in Medicine  
Date Published: 2013-04-01
Start Page: 041709
Language: English
PROVIDER: scopus
PUBMED: 23556879
DOI: 10.1118/1.4794487
DOI/URL:
Notes: --- - "Export Date: 1 May 2013" - "CODEN: MPHYA" - ":doi 10.1118/1.4794487" - "Source: Scopus"
Altmetric
Citation Impact
BMJ Impact Analytics
MSK Authors
  1. Marco Zaider
    171 Zaider