Multi-institutional comparison of whole slide digital imaging and optical microscopy for interpretation of hematoxylin-eosin-stained breast tissue sections Journal Article


Authors: Krishnamurthy, S.; Mathews, K.; McClure, S.; Murray, M.; Gilcrease, M.; Albarracin, C.; Spinosa, J.; Chang, B.; Ho, J.; Holt, J.; Cohen, A.; Giri, D.; Garg, K.; Bassett, R. L.; Liang, K.
Article Title: Multi-institutional comparison of whole slide digital imaging and optical microscopy for interpretation of hematoxylin-eosin-stained breast tissue sections
Abstract: Context.-Whole slide imaging (WSI) is now used for educational purposes, for consultation, and for archiving and quantitation of immunostains. However, it is not routinely used for the primary diagnosis of hematoxylineosin-stained tissue sections. Objective.-To compare WSI using the Aperio digital pathology system (Aperio Technologies, Inc, Vista, California) with optical microscopy (OM) for the interpretation of hematoxylin-eosin-stained tissue sections of breast lesions. Design.-The study was conducted at 3 clinical sites; 3 breast pathologists interpreted 150 hematoxylin-eosinstained slides at each site, 3 times each by WSI and 3 times each by OM. For WSI, slides were scanned using an Aperio ScanScope and interpreted on a computer monitor using Aperio ImageScope software and Aperio Spectrum data management software. Pathologic interpretations were recorded using the College of American Pathologists breast checklist. WSI diagnoses were compared with OM diagnoses for accuracy, precision (interpathologist variation), and reproducibility (intrapathologist variation). Results were considered accurate only if the interpretation matched exactly between WSI and OM. The proportion of accurate results reported by each pathologist was expressed as a percentage for the comparison of the 2 platforms. Results.-The accuracy of WSI for classifying lesions as not carcinoma or as noninvasive (ductal or lobular) or invasive (ductal, lobular, or other) carcinoma was 90.5%. The accuracy of OM was 92.1%. The precision and reproducibility of WSI and OM were determined on the basis of pairwise comparisons (3 comparisons for each slide, resulting in 36 possible comparisons). The overall precision of WSI was 90.5% in comparison with 92.1% for OM; reproducibility of WSI was 91.6% in comparison with 94.5% for OM, respectively. Conclusions.-In this study, we demonstrated that WSI and OM have similar accuracy, precision, and reproducibility for interpreting hematoxylin-eosin-stained breast tissue sections. Further clinical studies using routine surgical pathology specimens would be useful to confirm these findings and facilitate the incorporation of WSI into diagnostic practice.
Keywords: pathology; diagnosis; validation; design; interobserver; network; implementation; virtual microscopy; telepathology; prostate needle biopsies
Journal Title: Archives of Pathology & Laboratory Medicine
Volume: 137
Issue: 12
ISSN: 0003-9985
Publisher: College of American Pathologists  
Date Published: 2013-12-01
Start Page: 1733
End Page: 1739
Language: English
ACCESSION: WOS:000328263900004
DOI: 10.5858/arpa.2012-0437-OA
PROVIDER: wos
PUBMED: 23947655
Notes: Article -- Source: Wos
Altmetric
Citation Impact
BMJ Impact Analytics
MSK Authors
  1. Melissa P Murray
    123 Murray
  2. Dilip D Giri
    184 Giri