Performance of computer-assisted continuous infusion at low concentrations of intravenous sedatives Journal Article


Authors: Veselis, R. A.; Glass, P.; Dnistrian, A.; Reinsel, R.
Article Title: Performance of computer-assisted continuous infusion at low concentrations of intravenous sedatives
Abstract: We studied the performance of a target-controlled drug infusion device, computer-assisted continuous infusion (CACI). Forty-one volunteers received one of midazolam (n = 11), propofol (n = 10), thiopental (n = 10), or fentanyl (n = 10) in sedative concentrations. Concentrations were kept constant for 45-70 min at five sequential target concentrations in each subject. Twenty-six subjects had arterial sampling and 15 had venous sampling to determine drug concentrations. Median performance errors, median absolute performance error (MDAPE), wobble, divergence, and median absolute constancy error (MDACE), defined as error around mean actual concentration at each target, were calculated. CACI demonstrated significant performance errors, which were different among drugs. MDAPE (50%95% confidence interval) ranged from 22.9% (12.1%-39.6%) for propofol to 82.2% (36.0%-183.0%) for midazolam. Although performance errors could be large, CACI was able to maintain a constant serum concentration over time very successfully. The MDACE ranged from 5.6% (3.9%-17.3%) for fentanyl to 11.2% (8.9%-20.4%) for propofol. Few differences occurred between arterial and venous sampling, although when they occurred, arterial samples indicated larger errors. It is concluded that CACI is very successful at maintaining constant serum concentrations of these drugs at sedative concentrations. Arterial sampling should be used when the performance characteristics of an infusion device are being tested. However, venous sampling may be adequate to determine serum concentrations when a pseudo-steady state has been achieved.
Keywords: adult; clinical article; controlled study; clinical trial; controlled clinical trial; steady state; randomized controlled trial; midazolam; propofol; sedation; continuous infusion; statistical analysis; blood sampling; data analysis; intermethod comparison; drug blood level; double-blind method; infusions, intravenous; fentanyl; hypnotics and sedatives; calculation; thiopental; arterial blood; sedative agent; drug therapy, computer-assisted; venous blood; anesthesia induction; infusion pumps; humans; human; male; female; priority journal; article
Journal Title: Anesthesia and Analgesia
Volume: 84
Issue: 5
ISSN: 0003-2999
Publisher: Lippincott Williams & Wilkins  
Date Published: 1997-05-01
Start Page: 1049
End Page: 1057
Language: English
PUBMED: 9141930
PROVIDER: scopus
DOI: 10.1097/00000539-199705000-00018
DOI/URL:
Notes: Article -- Export Date: 17 March 2017 -- Source: Scopus
Altmetric
Citation Impact
BMJ Impact Analytics
MSK Authors
  1. Robert A Veselis
    98 Veselis
  2. Ruth A Reinsel
    78 Reinsel