Three-dimensional dosimetry imaging of I-125 plaque for eye cancer treatment Journal Article


Authors: Weaver, M.; Green, J.; Petasecca, M.; Lerch, M. L. F.; Cutajar, D.; Franklin, D.; Jakubek, J.; Carolan, M. G.; Conway, M.; Pospisil, S.; Kron, T.; Metcalfe, P.; Zaider, M.; Rosenfeld, A. B.
Article Title: Three-dimensional dosimetry imaging of I-125 plaque for eye cancer treatment
Abstract: Treatment of ocular cancers using eye plaque brachytherapy is now an established medical procedure. However, current QA for these eye plaques is quite rudimentary, limiting the opportunities for precise pre-tumour plaque customisation. This paper proposes and experimentally validates a new technique for imaging of eye plaque dose distributions using a high-resolution pixelated silicon detector. Results are presented demonstrating the 2D and 3D isodose surfaces produced using experimental data collected using this method. © 2010 Elsevier B.V. All rights reserved.
Keywords: three dimensional; radiotherapy; oncology; medical imaging; dosimetry; brachytherapy; high resolution; diseases; dose distributions; dosimetric imaging; ocular cancer; pixelated detector; customisation; experimental data; medical procedures; pixelated detectors; plaque brachytherapy; two dimensional
Journal Title: Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Volume: 633
Issue: Suppl. 1
ISSN: 0168-9002
Publisher: Elsevier Science BV  
Date Published: 2011-05-01
Start Page: S276
End Page: S278
Language: English
DOI: 10.1016/j.nima.2010.06.187
PROVIDER: scopus
DOI/URL:
Notes: --- - "Export Date: 17 August 2011" - "CODEN: NIMAE" - "Source: Scopus"