Noninvasive cardiac output monitoring with bioreactance as an alternative to invasive instrumentation for preclinical drug evaluation in beagles Journal Article


Authors: Heerdt, P. M.; Wagner, C. L.; DeMais, M.; Savarese, J. J.
Article Title: Noninvasive cardiac output monitoring with bioreactance as an alternative to invasive instrumentation for preclinical drug evaluation in beagles
Abstract: Introduction: Non-invasive measurement of cardiac output (CO) using bioreactance signals (NICOM) was recently validated in swine and human adults. The present study was designed to test the hypothesis that NICOM flow measurements can accurately measure CO and detect acute drug-induced changes in aortic blood flow (AoQ) ≥ 10% in beagles. Methods: Data from 5 anesthetized, open chest beagles used for preclinical screening of novel neuromuscular blocking drugs were analyzed for the study. AoQ was measured beat-to-beat with a probe on the aortic root and averaged over 30. s intervals. NICOM CO measurements were simultaneously obtained every 30. s. NICOM precision (random variation) and accuracy relative to AoQ were assessed from individual segments of steady-state data. The ability of NICOM to detect acute alterations in AoQ ≥ 10% was determined from other segments with dynamic change. Results: 516 simultaneous CO measurements between 826 and 2436. mL/min were analyzed. Steady-state measurements (20% of the dataset) revealed an average AoQ - NICOM difference (bias) of 63 ± 38. mL/min, a percent error of 6.1%, and a NICOM precision of 6.1%. Within the acute change dataset, 21/23 events reflecting a ≥ 10% change in beat-to-beat AoQ were detected by the NICOM (sensitivity of 0.91). In none of 10 instances where drug or fluid injection altered AoQ < 10% did the NICOM indicate a change (specificity of 1.0). Discussion: Continuous, non-invasive measurement of CO by bioreactance provides data that satisfactorily approximates invasive measurement of aortic blood flow in beagles. In addition, despite a 30. s interval between measurements, the NICOM appears to have sufficient fidelity to detect and quantify acute, drug-induced changes in CO. These data suggest that the NICOM may represent an alternative to open chest instrumentation for CO measurement in preclinical drug evaluation studies. © 2011 Elsevier Inc.
Keywords: controlled study; nonhuman; drug megadose; sensitivity and specificity; accuracy; animals; steady state; animal experiment; animal model; drug evaluation, preclinical; retrospective study; time factors; dog; dogs; heart output; heart stroke volume; cardiac output; non invasive measurement; drug induced disease; heart beat; suidae; muscle relaxant agent; toxicology; neuromuscular blocking agents; aorta; monitoring, physiologic; beagle; bioreactance; aorta flow; heart output measurement; physical phenomena
Journal Title: Journal of Pharmacological and Toxicological Methods
Volume: 64
Issue: 2
ISSN: 1056-8719
Publisher: Elsevier Inc.  
Date Published: 2011-09-01
Start Page: 111
End Page: 118
Language: English
DOI: 10.1016/j.vascn.2011.03.006
PROVIDER: scopus
PUBMED: 21440649
DOI/URL:
Notes: --- - "Export Date: 9 December 2011" - "CODEN: JPTME" - "Source: Scopus"
Altmetric
Citation Impact
BMJ Impact Analytics
MSK Authors
  1. Paul Heerdt
    46 Heerdt