Quantitative imaging of 1-124 using positron emission tomography with applications to radioimmunodiagnosis and radioimmunotherapy Journal Article


Authors: Pentlow, K. S.; Graham, M. C.; Lambrecht, R. M.; Cheung, N. K. V.; Larson, S. M.
Article Title: Quantitative imaging of 1-124 using positron emission tomography with applications to radioimmunodiagnosis and radioimmunotherapy
Abstract: Positron emission tomography (PET) is potentially useful for the quantitative imaging of radiolabeled antibodies, leading in turn to improved dosimetry in radioimmunotherapy. Iodine-124 is a positron-emitting nuclide with appropriate chemical properties and half-life (4.2 days) for such studies since the radiolabeling of antibodies with iodine is well understood and the half-life permits measurements over several days. Unfortunately, 1–124 has a complex decay scheme with many high-energy gamma rays and a positron abundance of only 25%. It has therefore been largely ignored as a PET-imaging nuclide. However, measurements made with phantoms and animals under realistic conditions using a BGO-based PET scanner have shown that satisfactory imaging and quantitation can be achieved. Investigations of spatial resolution, the linearity of regional observed count rate versus activity in the presence of other activity, and the visualization and quantitation of activity in spheres with different surrounding background activities were carried out with phantoms up to 22 cm in diameter. Compared with F-18, spatial resolution was only slightly degraded (13.5 mm FWHM vs 12 mm FWHM) while linearity was the same over a 10:1 activity range (0.015 to 0.15 MBq/ml for 1–124). The visualization and quantitation of spheres was also slightly degraded when using similar imaging times. Increasing the imaging time for 1–124 reduced the difference. To verify that the technique would work in vivo, measurements were made of human neuroblastoma tumors in rats which had been injected with 1–124 labeled 3F8 antibody. Although the number of samples was small, good agreement was achieved between image-based measurements and direct measurements of excised 4-g tumors. Thus quantitative imaging of 1–124 labeled antibodies appears to be possible under realistic conditions. © 1991, American Association of Physicists in Medicine. All rights reserved.
Keywords: controlled study; human tissue; unclassified drug; positron emission tomography; animal; animal experiment; antibodies, monoclonal; antigens, neoplasm; iodine radioisotopes; neuroblastoma; dosimetry; rat; transplantation, heterologous; quantitative imaging; iodine 124; rats; radioimmunotherapy; positron emission tomography (pet); tomography, emission-computed; quantitative diagnosis; cancer graft; immunoscintigraphy; iodine-124; human; priority journal; article; support, u.s. gov't, non-p.h.s.; models, structural; monoclonal antibody dosimetry; antibody 3f8
Journal Title: Medical Physics
Volume: 18
Issue: 3
ISSN: 0094-2405
Publisher: American Association of Physicists in Medicine  
Date Published: 1991-05-01
Start Page: 357
End Page: 366
Language: English
DOI: 10.1118/1.596728
PUBMED: 1870476
PROVIDER: scopus
DOI/URL:
Notes: Source: Scopus