Data from the analytical performance of the Abaxis Piccolo Xpress point of care analyzer in whole blood, serum, and plasma Journal Article


Authors: Murata, K.; Glaser, L.; Nardiello, M.; Ramanathan, L. V.; Carlow, D. C.
Article Title: Data from the analytical performance of the Abaxis Piccolo Xpress point of care analyzer in whole blood, serum, and plasma
Abstract: The objective of this study was to examine the analytical performance of 14 comprehensive metabolic panel analytes on the Abaxis Piccolo Xpress® Point of Care analyzer in serum, plasma, and whole blood. A method comparison was performed on all three specimen types intended for use on the Piccolo Xpress®: serum, heparinized plasma, and whole blood. This data is also presented in Murata et al. (2015) [1]. This article includes the actual Bland-Altman bias plots of the difference in results obtained for analytes in the comprehensive metabolic panel from the Abaxis Piccolo Xpress and the comparison instrument, the Ortho Vitros. © 2017
Keywords: clinical chemistry; point-of care testing
Journal Title: Data in Brief
Volume: 16
ISSN: 2352-3409
Publisher: Elsevier B.V.  
Date Published: 2018-02-01
Start Page: 81
End Page: 89
Language: English
DOI: 10.1016/j.dib.2017.11.006
PROVIDER: scopus
PMCID: PMC5694958
PUBMED: 29188225
DOI/URL:
Notes: Article -- Export Date: 2 January 2018 -- Source: Scopus