Authors: | Herr, H. W.; Reuter, V. E. |
Article Title: | Evaluation of new resectoscope loop for transurethral resection of bladder tumors |
Abstract: | Purpose: We evaluate a new resectoscope loop for transurethral resection of bladder tumors. Materials and Methods: Of 251 transurethral resections in 226 patients 111 were done with a conventional loop and 140 with the Olympus prototype model A2186 resectoscope loop. The quality of specimens provided for histological analysis was compared. Results: Tissue orientation was preserved and cautery artifact was reduced with the new loop compared to the standard resectoscope loop. Conclusions: Compared to a conventional resectoscope loop, the new loop resulted in a better tumor resection and helped facilitate histological evaluation of the tissue specimens. |
Keywords: | controlled study; major clinical study; bladder tumor; urinary bladder neoplasms; bladder neoplasms; transurethral resection; electrosurgery; electrode; endoscopes; humans; human; priority journal; article; cautery, urethra |
Journal Title: | Journal of Urology |
Volume: | 159 |
Issue: | 6 |
ISSN: | 0022-5347 |
Publisher: | Elsevier Science, Inc. |
Date Published: | 1998-01-01 |
Start Page: | 2067 |
End Page: | 2068 |
Language: | English |
PUBMED: | 9598520 |
PROVIDER: | scopus |
DOI: | 10.1016/s0022-5347(01)63249-x |
DOI/URL: | |
Notes: | Article -- Export Date: 12 December 2016 -- Source: Scopus |