Evaluation of tumor localization in respiration motion-corrected cone-beam CT: Prospective study in lung Journal Article


Authors: Dzyubak, O.; Kincaid, R.; Hertanto, A.; Hu, Y. C.; Pham, H.; Rimner, A.; Yorke, E.; Zhang, Q.; Mageras, G. S.
Article Title: Evaluation of tumor localization in respiration motion-corrected cone-beam CT: Prospective study in lung
Abstract: p Conclusions: Motion-corrected CBCT improves lung tumor localization accuracy and reduces motion artifacts in nearly all cases. Motion correction at end expiration using RCCT acquired at simulation yields similar results to that using a RCCT on the treatment day (2-3 weeks after simulation). C 2014 American Association of Physicists in Medicine.
Keywords: lung cancer; cone-beam computed tomography; image-guided radiation treatment; organ motion
Journal Title: Medical Physics
Volume: 41
Issue: 10
ISSN: 0094-2405
Publisher: American Association of Physicists in Medicine  
Date Published: 2014-10-01
Start Page: 101918-1
End Page: 101918-10
Language: English
DOI: 10.1118/1.4896101
PROVIDER: scopus
PUBMED: 25281970
PMCID: PMC4281114
DOI/URL:
Notes: Export Date: 3 November 2014 -- Source: Scopus