A rapid communication from the AAPM Task Group 201: Recommendations for the QA of external beam radiotherapy data transfer. AAPM TG 201: Quality assurance of external beam radiotherapy data transfer Journal Article


Authors: Siochi, R. A.; Balter, P.; Bloch, C. D.; Santanam, L.; Blodgett, K.; Curran, B. H.; Engelsman, M.; Feng, W.; Mechalakos, J.; Pavord, D.; Simon, T.; Sutlief, S.; Ronald Zhu, X.
Article Title: A rapid communication from the AAPM Task Group 201: Recommendations for the QA of external beam radiotherapy data transfer. AAPM TG 201: Quality assurance of external beam radiotherapy data transfer
Abstract: The transfer of radiation therapy data among the various subsystems required for external beam treatments is subject to error. Hence, the establishment and management of a data transfer quality assurance program is strongly recommended. It should cover the QA of data transfers of patient specific treatments, imaging data, manually handled data and historical treatment records. QA of the database state (logical consistency and information integrity) is also addressed to ensure that accurate data are transferred.
Keywords: treatment planning; quality assurance; data transfer; dicom; record & verify
Journal Title: Journal of Applied Clinical Medical Physics
Volume: 12
Issue: 1
ISSN: 1526-9914
Publisher: American College of Medical Physics  
Date Published: 2011-01-01
Start Page: 170
End Page: 181
Language: English
PROVIDER: scopus
PUBMED: 21330992
PMCID: PMC5718574
DOI/URL:
Notes: --- - "Export Date: 4 March 2011" - "Source: Scopus"