Quality assurance for image-guided radiation therapy utilizing CT-based technologies: A report of the AAPM TG-179 Journal Article


Authors: Bissonnette, J. P.; Balter, P. A.; Dong, L.; Langen, K. M.; Lovelock, D. M.; Miften, M.; Moseley, D. J.; Pouliot, J.; Sonke, J. J.; Yoo, S.
Article Title: Quality assurance for image-guided radiation therapy utilizing CT-based technologies: A report of the AAPM TG-179
Abstract: Purpose: Commercial CT-based image-guided radiotherapy (IGRT) systems allow widespread management of geometric variations in patient setup and internal organ motion. This document provides consensus recommendations for quality assurance protocols that ensure patient safety and patient treatment fidelity for such systems. Methods: The AAPM TG-179 reviews clinical implementation and quality assurance aspects for commercially available CT-based IGRT, each with their unique capabilities and underlying physics. The systems described are kilovolt and megavolt cone-beam CT, fan-beam MVCT, and CT-on-rails. A summary of the literature describing current clinical usage is also provided. Results: This report proposes a generic quality assurance program for CT-based IGRT systems in an effort to provide a vendor-independent program for clinical users. Published data from long-term, repeated quality control tests form the basis of the proposed test frequencies and tolerances. Conclusion: A program for quality control of CT-based image-guidance systems has been produced, with focus on geometry, image quality, image dose, system operation, and safety. Agreement and clarification with respect to reports from the AAPM TG-101, TG-104, TG-142, and TG-148 has been addressed. © 2012 American Association of Physicists in Medicine.
Keywords: quality assurance; cone-beam ct; fan-beam mvct; ct-on-rails
Journal Title: Medical Physics
Volume: 39
Issue: 4
ISSN: 0094-2405
Publisher: American Association of Physicists in Medicine  
Date Published: 2012-04-01
Start Page: 1946
End Page: 1963
Language: English
DOI: 10.1118/1.3690466
PROVIDER: scopus
PUBMED: 22482616
DOI/URL:
Notes: --- - "Export Date: 2 July 2012" - "CODEN: MPHYA" - "Source: Scopus"
Altmetric
Citation Impact
BMJ Impact Analytics