Article Title: | A phantom-based evaluation of a real-time tracking micro MLC delivery |
Abstract: | A cutting edge real-time tracking delivery system was evaluated for respiratory-correlated radiation therapy. The performed evaluations included Multi-Leaf Collimator (MLC) leaf position accuracy, MLC motion time delay, dosimetric performance, and tracking failure. The experiment-based results indicated that the dynamic MLC was capable of achieving the position accuracy in 1~2 mm and the MLC time delay in the range of 100 µs. The dosimetric comparison evidenced the significant advantage when utilising real-time tracking. The analysis of experiments also indicated the potential fail of tracking due to the impact of patient breathing patterns and the visibility of tumour. |
Keywords: | lung cancer; radiation therapy; dose distribution; radiotherapy treatment; breathing phantom; tumour motion; dual-layer micro MLC; intra-fractional motion; real-time tracking; multi-leaf collimator; MLC leaf position accuracy; MLC motion time delay; dosimetric performance; tracking failure; patient breathing patterns; tumour visibility; respiratory phantom |