A phantom-based evaluation of a real-time tracking micro MLC delivery Journal Article


Authors: Liu, Y.; Su, F. C.; Szegedi, M.; Lin, L.; Shi, C.; Papanikolaou, N.
Editor: Song, Y.
Article Title: A phantom-based evaluation of a real-time tracking micro MLC delivery
Abstract: A cutting edge real-time tracking delivery system was evaluated for respiratory-correlated radiation therapy. The performed evaluations included Multi-Leaf Collimator (MLC) leaf position accuracy, MLC motion time delay, dosimetric performance, and tracking failure. The experiment-based results indicated that the dynamic MLC was capable of achieving the position accuracy in 1~2 mm and the MLC time delay in the range of 100 µs. The dosimetric comparison evidenced the significant advantage when utilising real-time tracking. The analysis of experiments also indicated the potential fail of tracking due to the impact of patient breathing patterns and the visibility of tumour.
Keywords: lung cancer; radiation therapy; dose distribution; radiotherapy treatment; breathing phantom; tumour motion; dual-layer micro MLC; intra-fractional motion; real-time tracking; multi-leaf collimator; MLC leaf position accuracy; MLC motion time delay; dosimetric performance; tracking failure; patient breathing patterns; tumour visibility; respiratory phantom
Journal Title: International Journal of Biomedical Engineering and Technology
Volume: 8
Issue: 2-3
ISSN: 1752-6418
Publisher: Inderscience Publishers  
Date Published: 2012-01-01
Start Page: 274
End Page: 297
Language: English
DOI: 10.1504/IJBET.2012.046091
PROVIDER: manual
Notes: - Special Issue on 4D Medical Imaging and 4D Radiation Therapy
Altmetric
Citation Impact
BMJ Impact Analytics
MSK Authors
  1. Yulin Song
    116 Song