Technical Note: Impact of impurities on Yttrium-90 glass microsphere activity quantitation Review


Authors: Beattie, B. J.; Kirov, A. S.; Kesner, A. L.
Review Title: Technical Note: Impact of impurities on Yttrium-90 glass microsphere activity quantitation
Abstract: Background: Glass 90Y microspheres are produced with known radionuclide impurities. These impurities are not independently monitored. Clinical instruments, including ionization chamber dose calibrators and positron emmission tomography (PET) cameras, can be much more sensitive in detecting signals from these impurities than to signals from 90Y itself. Purpose: The “typical” levels of 90Y impurities have been studied to assess their impact on dosimetry during internal implantation, and for the management of waste. However, unaccounted-for decay spectra of impurities can also have an impact on dose calibrator and PET readings. Thus, even what might be considered negligibly small impurity fractions, can in principle cause substantial overestimates of the amount of 90Y activity present in a sample. To our knowledge, quantitative effects of radionuclide impurities in glass microspheres on activity measurements have not been documented in the field. As activity quantitation for dosimetry and its correlations with outcome becomes more prevalent, the effects of impurities on measurements may remain unaccounted for in dosimetry studies. Methods: In this letter, we review theoretical and physical considerations that will result in asymmetric errors in quantitation from 90Y impurities and estimate their typical and potential impact on clinical utilization. Among the common impurities 88Y is of particular concern for its impact on 90Y dose measurements because of its decay characteristics, along with other isotopes 91Y and 46Sc which can also impact measurements. Results: The typical level of 88Y impurities reported by the manufacturer should only cause small errors in dose calibrator and PET measurements made within the 12-day label-specified use-by period, up to 2.0% and 1.6%, respectively. However, the product specification max allowable impurity levels, specified by the manufacturer, leave open the potential for much greater bias from within the 12-day use-by period, potentially as high as 13.2% for dose calibrator measurements and 10.6% for PET from the 88Y impurities. Conclusions: While typical levels of impurities appear to have acceptable impact on patient absorbed dose, it should be noted that they can have adverse effects on 90Y radioactivity measurements. Furthermore, there is currently minimal independent verification and/or monitoring of impurity levels within the field. © 2023 American Association of Physicists in Medicine.
Keywords: adult; liver neoplasms; positron emission tomography; dosimetry; liver tumor; computer assisted emission tomography; radiometry; glass; radioisotopes; yttrium; embolization, therapeutic; yttrium radioisotopes; tomography, emission-computed; waste management; microsphere; yttrium 90; procedures; microspheres; yttrium-90; humans; human; article; drug impurity; artificial embolization; glass microspheres; technical notes; radionuclide impurities; yttrium-90 microspheres; yttrium-90 quantification; activity quantitation; impurity level; management of wastes; radionuclide impurity; yttrium-90 microsphere
Journal Title: Medical Physics
Volume: 51
Issue: 3
ISSN: 0094-2405
Publisher: American Association of Physicists in Medicine  
Date Published: 2024-03-01
Start Page: 2306
End Page: 2310
Language: English
DOI: 10.1002/mp.16828
PUBMED: 37956259
PROVIDER: scopus
DOI/URL:
Notes: Article -- MSK Cancer Center Support Grant (P30 CA008748) acknowledged in PubMed and PDF -- MSK corresponding author is Adam Kesner -- Source: Scopus
Altmetric
Citation Impact
BMJ Impact Analytics
MSK Authors
  1. Assen Kirov
    89 Kirov
  2. Bradley Beattie
    131 Beattie
  3. Adam Leon Kesner
    68 Kesner