Confounders of ultrasound attenuation imaging in a linear probe using the Canon Aplio i800 system: A phantom study Journal Article


Authors: Hänni, O.; Ruby, L.; Paverd, C.; Frauenfelder, T.; Rominger, M. B.; Martin, A.
Article Title: Confounders of ultrasound attenuation imaging in a linear probe using the Canon Aplio i800 system: A phantom study
Abstract: There have been studies showing attenuation imaging (ATI) with ultrasound as an approach to diagnose liver diseases such as steatosis or cirrhosis. So far, this technique has only been used on a convex probe. The goal of the study was to investigate the feasibility of ATI measurements using the linear array on a canon Aplio i800 scanner on certified phantoms. Three certified liver tissue attenuation phantoms were measured in five different positions using a linear probe. The effects of positioning and depth were explored and compared. The values were compared to the certified expected value for each phantom as well as the different measurement values for each measurement position. The ATI measurements on phantoms showed significant effect for the different probe positions and region of interest (ROI) depths. Values taken in the center with the probe perpendicular to the phantom were closest to certified values. Median values at 2.5–4.5 cm depth for phantoms 1 and 2 and 0.5–2.5 cm for phantom 3 were comparable with certified values. Measurements taken at a depth greater than 6 cm in any position were the least representative of the certified values (p-value < 0.01) and had the widest range throughout the different sessions. ATI measurements can be performed with the linear probe in phantoms; however, careful consideration should be given to depth dependency, as it can significantly affect measurement values. Remaining measurements at various depths within the 0.5–6.0 cm range showed deviation from the certified values of approximately 25%. © 2024 by the authors.
Keywords: ultrasound; liver; phantom; fat; attenuation
Journal Title: Diagnostics
Volume: 14
Issue: 3
ISSN: 2075-4418
Publisher: MDPI  
Date Published: 2024-02-01
Start Page: 271
Language: English
DOI: 10.3390/diagnostics14030271
PROVIDER: scopus
PMCID: PMC10855333
PUBMED: 38337786
DOI/URL:
Notes: Article -- Source: Scopus
Altmetric
Citation Impact
BMJ Impact Analytics
MSK Authors
  1. Lisa Ruby
    5 Ruby