Retrospective analysis of reduced energy switching and room switching times on throughput efficiency of a multi-room proton therapy center Journal Article


Authors: Mah, D.; Chen, C. C.; Nawaz, A. O.; Galbreath, G.; Shmulenson, R.; Lee, N.; Chon, B.
Article Title: Retrospective analysis of reduced energy switching and room switching times on throughput efficiency of a multi-room proton therapy center
Abstract: OBJECTIVE: To quantify how a control software upgrade changed beam delivery times and impacted efficiency and capacity of a multiroom proton therapy center. METHODS: A four-room center treating approximately 90 patients/day, treating for approximately 7 years with optimized operations, underwent a software upgrade which reduced room and energy switching times from approximately 30 to 20 s and approximately 4 s to ~0.5 s, respectively. The center uses radio-frequency identification data to track patient treatments and has software which links this to beam delivery data extracted from the treatment log server. Two 4-month periods, with comparable patient volume, representing periods before and after the software change, were retrospectively analyzed. RESULTS: A total of 16,168 and 17,102 fields were analyzed. For bilateral head and neck and prostate patients, the beam waiting time was reduced by nearly a factor of 3 and the beam delivery times were reduced by nearly a factor of 2.5. Room switching times were reduced more modestly. Gantry capacity has increased from approximately 30 patients to 40-45 patients in a 16-h daily operation. CONCLUSIONS: Many proton centers are striving for increased efficiencies. We demonstrated that reductions in energy and room switching time can significantly increase center capacity. Greater potential for further gains would come from improvements in setup and imaging efficiency. ADVANCES IN KNOWLEDGE: This paper provides detailed measured data on the effect on treatment times resulting from reducing energy and room switching times under controlled conditions. It helps validate the models of previous investigations to establish treatment capacity of a proton therapy center.
Keywords: retrospective studies; organization and management; retrospective study; time factors; cancer center; cancer care facilities; software; hospital management; efficiency, organizational; proton therapy; time factor; devices; capacity building; time to treatment; time-to-treatment; humans; human; health facility administration
Journal Title: British Journal of Radiology
Volume: 93
Issue: 1107
ISSN: 0007-1285
Publisher: British Institute of Radiology  
Date Published: 2020-01-01
Start Page: 20190820
Language: English
DOI: 10.1259/bjr.20190820
PUBMED: 31746631
PROVIDER: scopus
PMCID: PMC7066962
DOI/URL:
Notes: Article -- Source: Scopus
Altmetric
Citation Impact
BMJ Impact Analytics
MSK Authors
  1. Nancy Y. Lee
    871 Lee