Diode dosimetry of (103)Pd model 200 seed in water phantom Journal Article


Authors: Fontenla, D. P.; Ahmad, M.; Chiu-Tsao, S. T.; Anderson, L. L.
Article Title: Diode dosimetry of (103)Pd model 200 seed in water phantom
Abstract: The relative dose distribution around the 103Pd model 200 implant seed was measured with a computerized data acquisition system employing a p-n junction silicon diode immersed in a water phantom. Data are acquired in polar coordinates by computer control of (1) the diode distance from the seed center and (2) the rotation angle of seed about a transverse axis. Transverse axis data are compared with thermoluminescent dosimeter (TLD) measurements and a Monte Carlo calculation by others. © 1994, American Association of Physicists in Medicine. All rights are reserved.
Keywords: cancer growth; comparative study; radiation dose; methodology; radiotherapy dosage; models, theoretical; instrumentation; brachytherapy; plant seed; radioisotope; water; radioisotopes; models, anatomic; diode; theoretical model; palladium; anisotropy; water phantom; monte carlo method; audiovisual equipment; interstitial radiation; human; priority journal; article; diode dosimetry; 103pd seed; 2-d anisotropy
Journal Title: Medical Physics
Volume: 21
Issue: 6
ISSN: 0094-2405
Publisher: American Association of Physicists in Medicine  
Date Published: 1994-06-01
Start Page: 817
End Page: 820
Language: English
DOI: 10.1118/1.597327
PUBMED: 7935219
PROVIDER: scopus
DOI/URL:
Notes: Source: Scopus